PD6001-6H5
Series: PD6001 ProVu Analog Input Feet & Inches Meter Brand: Precision Digital Corporation
Price
Request a Quote
Stock
Call for Pricing & Availablility
- Product - SunBright Display Model
- Power Supply - 85-265 Vac
- Output - 2 Relays & 4-20 mA
Specifications
Accuracy
- Isolated 4-20 mA Transmitter Output: ± 0.1 % of Span ± 0.004 mA
- Process Input (Square Root & Programmable Exponent Accuracy Range): 10 % to 100 % of Calibrated Span
- Process Input: ± 0.03 % of Calibrated Span ± 1 Count
Alarm
- Relay: High/Low
Approvals & Certifications
- CE
- UL (C,US)
- UL: E160849, 508 Industrial Control Equipment
Calibration
- Filter Bypass: Programmable from 0.1 to 99.9 % of Calibrated Span
- Process Input, Isolated 4-20 mA Transmitter Output: Factory Calibrated 4-20 mA
- Recalibration is Recommended at Least Every 12 Months
Communication
- Baud Rate: 300 to 19,200 bit/sec
- Byte-to-Byte Timeout: 0.01 to 2.54 seconds
- Compatibility: USB 2.0
- Data: 8-bit (1 Start Bit, 1 or 2 Stop Bit)
- Parity - Stop Bit: 1-bit, 2-bits (Even, Odd or None)
- Protocol: Modbus RTU
- RS-232 - PDA1232
- RS-422
- RS-485 - PDA1485
Current
- Required External Fuse: UL Recognized, 5 A, Max.
Deadband
- Relay: 0 to 100 % of Span, User Programmable
Delay Time
- 0 to 999.9 seconds (On & Off Relay)
- Communication (Transmit Time Delay): Programmable between 0 to 199 ms
- Turn Around Delay: < 2 ms (Fixed)
Display
- 6-Digit Display (-99999 to 999999)
- Dual-Line 6-Character
- Lower Display: 0.46" (12 mm) High
- Overrange & Underrange: Display Flashes -99999 to 999999
- Process Input (Low-Flow Cut-Off): 0-999,999 (0 Disables Cut-Off Function)
- Red LEDs
- Upper feet & inches Display: 0.60" (15 mm) High
- Upper: 0.60" High, Lower: 0.46" High
Electrical Connectors
- Relay: Form A (4 SPST)
- Relay: Form C (2 or 4 SPDT)
- Removable Screw Terminal Blocks Accept 12 to 22 AWG Wire
- RJ45 for External Relays, Digital I/O, & Serial Communication Adapters
Electrical Contacts
- Normally Open
Electrical Rating
- 1/14 hp (≈50 W) at 125/250 Vac for Inductive Loads Such as Contactors, Solenoids
- 3 A at 125/250 Vac Resistive Load
- 3 A at 30 Vdc Resistive Load
Enclosure Finish
- Black
Enclosure, Body Material
- Plastic
Environmental Protection
- Explosion-proof
- IP65
- IP66
- NEMA 4X
Function
- Process Input (Signal Input Conditioning): Linear
- Process Input (Signal Input Conditioning): Round Horizontal Tank Volume Calculation
Humidity
- 0 % to 90 % RH, Non-Condensing
Input
- ± 10 Vdc (0-5, 1-5, 0-10 V) Modbus PV (Slave)
- 0-5 V
- 1-5 V
- F4-Digital Input Contacts: 3.3 Vdc
- F4-Digital Input Logic: 0-1.25 Vdc (Low)
- F4-Digital Input Logic: 3-5 Vdc (High)
- Process Input (Field Selectable): 4-20 mA
- Process Input Overload: Current Input Protected by Resettable Fuse, 30 Vdc, Max.
Input Impedance
- Process Input at Current Range: 50 to 100 ohms (Depending on Resettable Fuse Impedance)
- Process Input at Voltage Range: > 1 MΩ
Isolation
- 4 kV Input/Output-to-Power-Line
- 500 V Input to Output or Output to P+ Supply
Linearization
- Exponential
- Process Input (Multi-Point Linearization): 2 to 32-Points for PV or PV1
- Process Input (Multi-Point Linearization): 2 to 8-Points for PV2 (Dual-Scale Level Feature)
Loop Resistance
- Isolated 4-20 mA Transmitter Output: 10 to 700 ohms at 24 Vdc Power Supply
- Isolated 4-20 mA Transmitter Output: 100 to 1,200 ohms at 35 Vdc (External) Power Supply
Mounting
- 1/8" DIN Panel Cutout Required: 3.622" x 1.772" (92 mm x 45 mm)
- DIN Rail Mounting Kit
- Two Panel Mount Bracket
Operating Temperature
- -40° to 65° C (-40° to 149° F)
Output
- Isolated 4-20 mA Transmitter Output
- Isolated 4-20 mA Transmitter Output (Analog Output Programming): 23 mA - Max.
- Isolated 4-20 mA Transmitter Output (Scaling Range): 1-23 mA for Any Display Range
Power Options
- 12-24 Vdc ± 10 %, 15 W, Max.
- 20 W, Max.
- 85-265 Vac, 50/60 Hz
- 90-265 Vdc
Power Supply
- 24 V at 200 mA
- Additional Power Supply: 24 V at 40 mA
- Isolated 4-20 mA Transmitter Output (External Loop Power Supply): 35 Vdc, Max.
- Isolated 4-20 mA Transmitter Output: 24 Vdc ± 10 % (Terminals I+ & R, P+ & P)
- Isolated Transmitter Power Supply: 10 Vdc at 50 mA
- Isolated Transmitter Power Supply: 12-24 Vdc at 100 mA - Max.
- Isolated Transmitter Power Supply: 5-10 Vdc at 50 mA
- Isolated Transmitter Power Supply: 85-265 Vac at 200 mA - Max.
Programming
- Digital Inputs
- Four Front Panel Buttons
- Modbus Registers
- PC & MeterView Pro Software
Relay Operation
- Automatic (Non-Latching)
- Latching (Requires Manual Acknowledge)
- Off (Disable Unused Relays & Enable Interlock Feature, Manual ON/OFF Control Mode)
- Pump Alternation Control (2 to 8 Relays)
- Sampling (Based on Time)
Storage Temperature
- -40° to 85° C (-40° to 185° F)
Supply Voltage
- Internally Selectable Jumper for 5-24 Vdc
Switch Type
- SPDT
System Requirements
- Driver: Windows 10 32/64-bit
- Driver: Windows 7 32/64-bit
- Driver: Windows 98/SE, ME, 2000, Server 2003/2008
- Driver: Windows Vista 32/64-bit
- Driver: Windows XP 32/64-bit
Thermal Drift
- Isolated 4-20 mA Transmitter Output: 0.4 μA/°C, Max.
- Isolated 4-20 mA Transmitter Output: 0.8 μA/°C, Max.
- Process Input: 0.005 % of Calibrated Span/°C, Max.
- Process Input: 0.01 % of Calibrated Span/°C, Max.
Torque
- Tightening Torque (Screw Terminal Connectors): 0.56 Nm (5 in/lb)
Update Time