TT301 Temperature Transmitter
Brand: Smar International CorpSpecifications
- Output: HART Protocol, 2-Wire 4-20mA
- Input Range: -200°C to +1750°C, -328°F to +3182°F
- Minimum Span: +10°C to +60°C, +50°F to +140°F
- Control: PID Capability
- Power Supply
- Bus Powered: 12 to 45 Vdc.
- Digital Accuracy: ±0.2 to ±0.6°C
- Display: 4-1/2 digit numerical and 5-Character alphanumerical LCD indicator
- Burnout: Upscale
- Mounting Bracket: Carbon Steel or 316 Stainless Steel
- Electrical Connection
- 1/2"-14 NPT
- PG 13.5 DIN
- M20 X 1.5
- Damping Adjustment: User Configurable from 0 to 32 seconds
- Power Supply Effect: ± 0.005% of calibrated span per volt
- CSA Intrinsically Safe, Explosion Proof, Dust Ignition Proof, Non-Incendive
- Order includes a SMAR Magnetic Screwdriver for adjustments if the LCD Display option is selected in the configuration below
Specifications
Accuracy
- ± 0.02 % of Span
Alarm
- HART: in Case of Sensor or Circuit Failure, the Self Diagnostics Drives the Output to 3.6 to 21.0 mA, According to the User's Choice & NAMUR NE43 Specification.
Ambient Temperature Range
- -40° to 85° C (-40° to 185° F)
Approvals & Certifications
- CEPEL
- CSA
- EXAM
- FM
- NEMKO ATEX
Burn-Out
- Upscale
Communication
- HART: 2-Wire, 4-20 mA According to NAMUR NE43 Specification, with Super-imposed Digital Communication.
Configuration
- Software CONF401, DDCON100, or HpC301 & HpC401. Can Also be Configured Using DD & FDT/DTM Tools, & Can be Partially Configured through Local Adjustment
Control Functions Characteristics
- HART: PID Control, Alarm & Set Point Ramp Generator
Damping Adjustment
- User Configurable from 0 to 32 seconds (via Digital Communication).
Digital Stability
- ± 0.1 % of Reading or +0.1° C (0.18° F), whichever is greater, for 1-Year for Thermocouple
- ± 0.1 % of Reading or +0.1° C (0.18° F), whichever is greater, for 2-Years for RTDs
Display
- 4-1/2-Digit LCD Indicator
Electrical Connectors
- 1/2" NPT
Electromagnetic Interference Effect
- Approved According to IEC 61326-1: 2006, IEC 61326-2-3: 2006, IEC 61000-6-4: 2006, IEC 61000-6-2: 2005
Enclosure Finish
- Gray Munsell 6.5 Polyester Paint
Enclosure, Body Material
- 316 Stainless Steel IP/TYPE, 316 Stainless Steel for Saline Atmosphere IPW/TYPEX
- Copper-Free Aluminum IPW / TYPEX, Aluminum for Saline Atmosphere IPW / TYPEX, Aluminum IP / TYPE
Humidity
- 0 to 100 % RH
Input Power
- HART: 12-45 Vdc
Mounting
- 316 Stainless Steel Bracket & Accessories
- Can be Attached Directly to the Sensor. with an Optional Bracket can be Installed on a 2" Pipe or Fixed on a Wall or Panel.
- Carbon Steel Bracket & Accessories
Operating Temperature - Display
- -20° to 80° C (-4° to 176° F)
- -40° to 85° C (-40° to 185° F), without Damage
Output
- 2-Wire 4-20 mA
Power Supply Effect
- ± 0.005 % of Calibrated Span Per Volt
Sensor, Probe Connection
- 2-Wire
- 3-Wire
- 4-Wire
- Backup
- Differential
Sensor, Probe Type
- J Thermocouple
- K Thermocouple
- RTD Pt100 - IEC
- R Thermocouple
- S Thermocouple
Span & Zero Adjustments
- Non-Interactive, Via Local Adjustment & Digital Communication
Storage Temperature
- -40° to 100° C (-40° to 212° F)
Temperature Effect
- mV (-10 to 100 mV) TC (NBS: E, J, K, N): ± 0.03 % of the Input mV or 0.01 mV whichever is greater (10° C Variation)
- mV (-6-22 mV), TC (NBS: B, R, S, T): ± 0.03 % of the Input mV or 0.002 mV whichever is greater -10° C Variation
- Ohms: 0 to 400 ohms, RTD (IEC: Pt100): ± 0.03 % of the Input Resistance or 0.04 ohms, whichever is greater