Accuracy
- Chart Speed Accuracy: ± 0.1 % (Expansion & Contraction of Paper is Not Included)
- Clock Accuracy: ± 50 ppm or Less (Monthly Error, About 2 minutes)
- Reference Junction Compensation (K, E, J, T, N, L, U, PN Thermocouple): ± 0.5° C (± 0.9° F)
- Reference Junction Compensation (R, S, B, W Thermocouple): ± 1° C (± 1.8° F)
- Thermocouple, Resistance Bulb, DC Voltage: ± (0.15 % + 1-Digit)
- Thermocouple, RTD, DC Voltage: ± 0.25 % of Recording Span
Communication
- Transmission Method: Half-Duplex Bit Serial, 500 Kbps Up to, Max.
Consumption
- 100 Vac & 24 Vdc with All Options Approximately: 37 VA
- 240 Vac with All Options Approximately: 56 VA
Electrical Rating
- 3 A at 240 Vac / 30 Vdc
- Contact Capacity: 0.05 A at 12 Vdc, Normally Open
Environmental Protection
- Front: IEC IP50
- Terminal: IP20
Function
- Alarm
- Recording Switch on Camera
Humidity
- Chart Paper Influence (Contraction): 35 % Relative Humidity, 0.5 %, Max.
- Chart Paper Influence (Expansion): 85 % Relative Humidity, 0.4 %, Max.
- Operating Environment: 20 to 80 % Relative Humidity, Non-Condensing
- Transportation/Storage Condition: 5 to 90 % Relative Humidity, Non-Condensing
Input
- DC Voltage Input (5 V, 50 V Range): -100 to 100 Vdc or Less
- Thermocouple, Resistance Bulb & DC Voltage (50 mV, 500 mV Range): -10 Vdc or Less
Input Signal
- -50 to 50 V
- -500 to 500 mV
Insulation Resistance
- 100 megaohms or more (Between Each Terminal & Earth, at 500 Vdc)
Materials of Construction
- Front Door Frame: Polycarbonate with Glass Wool
Mounting
- Inclined up to 30° Backwards from the Vertical: Panel
Mounting Position
- Front & Left/Right Inclination: 0°
- Rear Inclination: 30°
Mounting Position Effect
- Change in Indication: ± (0.1 % + 1-Digit) - Max.
- Changing in Recording: ± 0.2 % of Recording Span, Max.
Operating Temperature
- 0° to 50° C (32° to 122° F)
Output Noise
- Common Mode Noise Rejection: 120 dB at 50, 60 Hz ± 0.1 Hz
- Series Mode Noise Rejection: 30 dB at 50, 60 Hz ± 0.1 Hz
Power Supply
- 100-240 Vac (10 %, ± 15 %), 50/60 Hz
- 24 Vdc, ± 10 %
Power Supply Effect
- Change in Indication: ± (0.1 % + 1-Digit) - Max.
- Change in Recording: ± 0.2 % of Recording Span, Max.
Resistance
- Input (5 V & 50 V): ≥ 1 Megaohm
- Input (Thermocouple & 50 mV): > 10 megaohms
- Iput (500 mV): > 100 kiloohms
- Signal Source Resistance (Resistance Bulb Input): less than 10 ohms Per Line
- Signal Source Resistance (Thermocouple Input): 1 Kiloohm or Less Voltage
Resistance Effect
- Change in Indication (Thermocouple & RTD): ± (0.1 % + 1-Digit)
- Change in Recording (Thermocouple & RTD): ± 0.2 % of Span
- Voltage Input: 0.1 % Change of Resistance
Resolution
- DC Voltage: 10 µA to 10 mA
- Thermocouple, Resistance Bulb: 0.1° C
- Thermocouple, RTD, DC Voltage: 0.1 mm Min.
Sensor, Probe Type
- B Thermocouple as Per EN 60584
- C Thermocouple
- D Thermocouple
- E Thermocouple
- J Thermocouple
- K Thermocouple
- L Thermocouple as Per DIN 43710
- N Thermocouple
- PN Thermocouple
- RTD
- R Thermocouple
- S Thermocouple
- T Thermocouple
- U Thermocouple as Per DIN 43710
Shock
- 294 m/sec² (30 G) or Less
Storage Temperature
- -10° to 60° C (14° to 140° F)
Supply Voltage
- Withstand Voltage: between Alarm Terminals 750 Vac for 1 Minute
- Withstand Voltage: Input Terminal - Input Terminal, Input Terminal - Ground, Power Terminal - Input Terminal 500 Vac for 1 Minute
- Withstand Voltage: Power Supply Terminal - Ground 2,000 Vac for 1 Minute
Temperature Effect
- Change in Indication: ± (0.3 % + 1-Digit)/10° C (18° F) - Max.
- Change in Recording: ± 0.5 %/10° C (18° F) - Max.
Temperature Range
- RTD: -200° to 600° C (-328° to 1,112° F)
- Thermocouple: -200° to 1,760° C (-328° to 3,200° F)
Terminal Strip
- External Terminals: Screw Terminal Strip (M4 Screws)
Update Time
- Warm-Up Time: 30 minutes or more
Vibration
- Operating Environment: 10-60 Hz, 0.2 m/sec² (0.02 G) or Less
- Transportation/Storage Condition: 2.45 m/sec² (3 G) or Less
Vibration Effect
- Change in Indication: ± (0.1 % + 1-Digit) - Max.
- Change in Recording: ± 0.2 % of Recording Span, Max.